Exhibiting at SEMICON Japan 2026 (Dec 9–11)? We are now accepting advertising inquiries from exhibitors.Ads for SEMICON Japan 2026Learn more

Accelerated life test (HAST) equipment

EHS-432

A large highly accelerated stress test system (HAST chamber) from ESPEC with terminals for bias testing, compliant with IEC 60068-2-66. The test chamber volume is 130 L, or 180 L for the long type.

Maker: ESPEC · Official product page

ESPEC
  • Free to useQuotes and inquiries are free of charge
  • Phone number is optionalYou can send the form without it
Official product page ↗
How it works
  1. Fill in the form and send itEnter your company name, contact name, email address and request. Phone, quantity and timing are optional.
  2. Semirai forwards your requestSemirai receives your request and forwards it to ESPEC. Your details are shared with the maker only with your consent.
  3. The maker repliesReplies usually arrive within 2–3 business days.

Can't find a product, or unsure which to choose?

Free consultation

Specifications


TypeHAST (unsaturated control and wet saturated control; the M type adds wet/dry bulb temperature control for 3 modes)
Test chamber volume130 L (EHS-432(M)) / 180 L (long type EHS-432(M)-L)
Unsaturated control105.0–162.2°C, 75–100% rh, 0.0196–0.392 MPa (gauge)
StandardsIEC 60068-2-66 and others (IEC 60749-4, JESD22-A110E, etc. with M instrumentation)

Specifications as published by the maker. See the sources below.

Overview


EHS-432 is a large highly accelerated stress test system (HAST chamber) from ESPEC. It has terminals for bias testing under constant voltage and signal application, complies with the international standard IEC 60068-2-66 despite its large size, and can also run pressure cooker tests.

In addition to the standard type with unsaturated control and wet saturated control, there is an M type that adds wet/dry bulb temperature control for 3 modes. The test chamber volume is 130 L, and the long type EHS-432(M)-L, with extended depth, is 180 L. Under unsaturated control, the temperature range is 105.0–162.2°C and humidity is 75–100% rh.

To run tests compliant with the semiconductor standards IEC 60749-4, JEITA ED-4701, JESD22-A110E and JESD22-A102E, users select M instrumentation. The official site presents test examples such as wire bonding continuity evaluation of semiconductor packages (130°C/85% rh, 100 h).

Key features


  • Terminals for bias testing
  • Standard type (2 modes) and M type with wet/dry bulb temperature control (3 modes)
  • 130 L standard and 180 L long type
  • Mechanical door lock and door lock safety mechanism as standard

Applications


  • Wire bonding continuity evaluation of semiconductor packages
  • HAST of semiconductor devices based on IEC 60749-4, JESD22-A110E and others
  • Migration and whisker evaluation of printed circuit boards

FAQ


What procedures are needed to install EHS-432?

According to the official site, EHS-432 is a Class 1 pressure vessel under Japanese law. An installation notice and other procedures must be filed with the local Labor Standards Inspection Office at least 30 days before delivery and installation. A Class 1 pressure vessel operations chief (ordinary) must be appointed to use and manage it.

Other accelerated life test (hast) equipment


All accelerated life test (hast) equipment and Japanese suppliers

More from ESPEC


Sources


  1. ESPEC: large HAST chamber (highly accelerated stress test system) product page

More on this topic


Last updated: November 13, 2026