EHS-432
A large highly accelerated stress test system (HAST chamber) from ESPEC with terminals for bias testing, compliant with IEC 60068-2-66. The test chamber volume is 130 L, or 180 L for the long type.
Maker: ESPEC · Official product page
- Free to useQuotes and inquiries are free of charge
- Phone number is optionalYou can send the form without it
How it works
- Fill in the form and send itEnter your company name, contact name, email address and request. Phone, quantity and timing are optional.
- Semirai forwards your requestSemirai receives your request and forwards it to ESPEC. Your details are shared with the maker only with your consent.
- The maker repliesReplies usually arrive within 2–3 business days.
Can't find a product, or unsure which to choose?
Free consultationSpecifications
| Type | HAST (unsaturated control and wet saturated control; the M type adds wet/dry bulb temperature control for 3 modes) |
|---|---|
| Test chamber volume | 130 L (EHS-432(M)) / 180 L (long type EHS-432(M)-L) |
| Unsaturated control | 105.0–162.2°C, 75–100% rh, 0.0196–0.392 MPa (gauge) |
| Standards | IEC 60068-2-66 and others (IEC 60749-4, JESD22-A110E, etc. with M instrumentation) |
Specifications as published by the maker. See the sources below.
Overview
EHS-432 is a large highly accelerated stress test system (HAST chamber) from ESPEC. It has terminals for bias testing under constant voltage and signal application, complies with the international standard IEC 60068-2-66 despite its large size, and can also run pressure cooker tests.
In addition to the standard type with unsaturated control and wet saturated control, there is an M type that adds wet/dry bulb temperature control for 3 modes. The test chamber volume is 130 L, and the long type EHS-432(M)-L, with extended depth, is 180 L. Under unsaturated control, the temperature range is 105.0–162.2°C and humidity is 75–100% rh.
To run tests compliant with the semiconductor standards IEC 60749-4, JEITA ED-4701, JESD22-A110E and JESD22-A102E, users select M instrumentation. The official site presents test examples such as wire bonding continuity evaluation of semiconductor packages (130°C/85% rh, 100 h).
Key features
- Terminals for bias testing
- Standard type (2 modes) and M type with wet/dry bulb temperature control (3 modes)
- 130 L standard and 180 L long type
- Mechanical door lock and door lock safety mechanism as standard
Applications
- Wire bonding continuity evaluation of semiconductor packages
- HAST of semiconductor devices based on IEC 60749-4, JESD22-A110E and others
- Migration and whisker evaluation of printed circuit boards
FAQ
What procedures are needed to install EHS-432?
According to the official site, EHS-432 is a Class 1 pressure vessel under Japanese law. An installation notice and other procedures must be filed with the local Labor Standards Inspection Office at least 30 days before delivery and installation. A Class 1 pressure vessel operations chief (ordinary) must be appointed to use and manage it.
Other accelerated life test (hast) equipment
- AMM-C / AMM-W · ESPEC
- EHS series HAST chambers (standard type) · ESPEC
All accelerated life test (hast) equipment and Japanese suppliers
More from ESPEC
- Platinous J series · Temperature and humidity test chambers
- RBC series burn-in chambers · Burn-in systems
- RBS · Burn-in systems
- TCC-151W · Thermal shock test chambers
- TSA series thermal shock chambers · Thermal shock test chambers
- TSD-101-W · Thermal shock test chambers
- TSE-12-A · Thermal shock test chambers
Sources
Accelerated life test (HAST) equipment
Last updated: November 13, 2026