PN-12α
Napson's contact PN type tester using the thermoelectric method. A pen-type probe is placed on the sample to determine the conductivity type (PN) of single-crystal silicon wafers and ingots.
Maker: Napson · Official product page
- Free to useQuotes and inquiries are free of charge
- Phone number is optionalYou can send the form without it
How it works
- Fill in the form and send itEnter your company name, contact name, email address and request. Phone, quantity and timing are optional.
- Semirai forwards your requestSemirai receives your request and forwards it to Napson. Your details are shared with the maker only with your consent.
- The maker repliesReplies usually arrive within 2–3 business days.
Can't find a product, or unsure which to choose?
Free consultationSpecifications
| Type | Contact, thermoelectric method (pen-type probe with integrated hot/cold probes) |
|---|---|
| Measurement targets | Semiconductor and solar cell materials (silicon, polysilicon and others) |
| Sample size | 30 × 30 mm or larger |
| Resistivity range for PN testing | About 1 m to 20 k Ω·cm |
Specifications as published by the maker. See the sources below.
Overview
PN-12α is Napson's contact PN type tester using the thermoelectric method. A pen-type probe with integrated hot and cold probes is placed in contact with the sample to determine its conductivity type (PN).
It supports single-crystal silicon wafers, bulk and ingots, and handles samples of 30 × 30 mm or larger. The range of sample resistivity for PN testing is about 1 m to 20 k Ω·cm.
For samples that are hard to test with the integrated probe, a separate-type hot/cold probe is also available. Napson says some samples, such as polycrystalline silicon, wafers with films, and samples with mirror-finished or oxidized surfaces, may not be testable.
Key features
- Conductivity type (PN) testing by the thermoelectric method
- Pen-type probe with integrated hot/cold probes
- Separate-type probes also available
- Contact type with 1-point measurement
Applications
- Conductivity type testing of single-crystal silicon wafers, bulk and ingots
FAQ
Are there samples that PN-12α cannot test?
The official page says some samples, such as polycrystalline silicon, wafers with films, and samples with mirror-finished or oxidized surfaces, cannot be tested.
Terms explained
Light Point Defect (LPD) · Photoluminescence (PL) · X-ray Diffraction (XRD) · Raman Spectroscopy
Other wafer substrate inspection equipment
- EPM-2200 · Kobelco Research Institute
- LEP series · Kobelco Research Institute
- SBW series flatness and shape measurement systems · Kobelco Research Institute
- MAGICS M5640 · Lasertec
- EC-80 · Napson
- NC-6800FLA · Napson
- NC-80MAP · Napson
- PN-50α · Napson
All wafer substrate inspection equipment and Japanese suppliers
More from Napson
- FLA-200 · Wafer flatness and thickness metrology
- NC-2000FLA · Wafer flatness and thickness metrology
- NC-600MAP · Sheet resistance measurement
- PN-8LP · Wafer substrate inspection equipment
- RG-2000AL(RG-3000AL) · Sheet resistance measurement
- RT-3000 / RG-2000 · Sheet resistance measurement
- WS-3000 · Sheet resistance measurement
- WS-8800 · Sheet resistance measurement
Sources
Wafer substrate inspection equipment
Last updated: November 13, 2026