SEM vs TEM: what is the difference?

Both instruments use electrons instead of light to see features far below the optical diffraction limit, and both are essential to semiconductor manufacturing. They differ in how the beam interacts with the sample, and that difference decides what each one can show, what the sample must look like and where in the fab it is used.

The short answer


A scanning electron microscope (SEM) scans a focused electron beam across a sample and builds an image from the electrons that come back out of the surface, so it can look at bulk samples and even whole wafers at roughly nanometre resolution. A transmission electron microscope (TEM) sends a high-energy beam through a lamella thinned to around 100 nm or less and images the transmitted electrons, which resolves lattice planes and, with aberration correction, individual atomic columns. In a fab, SEMs measure and review on the line; TEMs are used off-line for process development and failure analysis, at the cost of destroying the sample.

Side-by-side comparison


AspectSEMTEM
Imaging principleA focused beam is rastered over the surface; secondary and backscattered electrons are collected point by pointA broad or focused beam passes through a thin specimen; transmitted and diffracted electrons are imaged by lenses
Information obtainedSurface topography, cross-section shape, compositional contrastInternal structure projected through the thickness, lattice images, interfaces, crystal defects
Typical resolutionAround 1 nm, depending on instrument and conditionsSub-nanometre lattice resolution; below 0.1 nm with aberration correction
Accelerating voltageFrom a few hundred volts to about 30 kVTypically 80-300 kV
Sample thicknessNo limit; whole wafers can be loaded in in-line toolsMust be electron-transparent, usually 100 nm or less
Sample preparationOften none for top-down imaging; cleaving or FIB milling for cross-sectionsThinning by FIB or other methods is mandatory and takes hours
Destructive?Low-voltage in-line SEMs return the wafer to the process flowYes; the area examined is cut out of the wafer
Field of viewCan zoom continuously from wide areas to high magnificationLimited to the small region captured in the lamella
Main fab usesCD-SEM linewidth metrology, defect review, cross-section imagingGate stack thickness and interfaces, GAA nanosheet checks, failure analysis
Common add-on analysisEDX for elemental compositionSTEM imaging, EDX and EELS for elemental and chemical mapping

Surface signals versus transmitted electrons


In an SEM, the beam is focused to a spot a few nanometres across and scanned in a raster pattern. Where it lands, secondary electrons and backscattered electrons leave the sample, and the detector signal at each point becomes one pixel. Secondary electrons escape only from a very shallow depth, so the image is dominated by surface topography; backscattered electrons add contrast from atomic number.

In a TEM, the specimen is made thin enough for electrons to pass through. Electrons scattered and diffracted by atoms inside the specimen are focused by electromagnetic lenses, so the image shows the internal structure projected through the thickness rather than the surface. Under suitable conditions, rows of atoms appear directly as lattice fringes.

Why the production line uses SEM


The electron microscopes found in high-volume fabs are SEM-based. A CD-SEM measures the width of resist and etched patterns from above without breaking the wafer, usually at a low landing energy of around 1 kV or less to limit charging and beam damage to resist. A defect review SEM drives to the coordinates reported by an optical or e-beam inspection tool, images each defect at high magnification and classifies it.

After either measurement the wafer continues through the process. A TEM cannot play this role, because preparing a TEM sample means cutting a piece out of the wafer.

When only TEM will do


Once gate dielectrics and GAA nanosheets are only a few nanometres thick, the edges of each layer blur in an SEM image and their thickness cannot be measured reliably. The thickness of a high-k film and its interfacial layer, the fill of a metal gate, or stacking faults in an epitaxial layer all need atomic-scale resolution.

Scanning TEM (STEM) mode, where a fine probe is scanned across the lamella, can be combined with EDX and EELS so that structure and elemental distribution are mapped at the same location. For new process development, and for pinning down the physical cause of an electrical failure, TEM is usually the final confirmation step.

Sample preparation is the bottleneck


In practice, the limiting factor for TEM is not the microscope but the specimen. Cutting a lamella that contains a specific transistor with a focused ion beam (FIB) and thinning it to a few tens of nanometres takes a skilled operator several hours, and the ion beam can leave a damaged surface layer that distorts the result.

Labs therefore work in sequence: survey a wide area in the SEM, narrow down the anomaly, and then prepare a TEM lamella at the exact site. A dual-beam FIB-SEM combines both steps in one chamber, so the operator can watch the cut with the electron beam while milling with the ion beam.

FAQ


Can an SEM image cross-sections?

Yes. A wafer can be cleaved, or a cross-section can be milled with a FIB, and the exposed face imaged in the SEM. For films only a few nanometres thick, or for the atomic arrangement at an interface, the resolution of a TEM is still needed.

Is STEM closer to SEM or to TEM?

STEM uses the same thin lamella as TEM and detects transmitted electrons, but it forms the image by scanning a focused probe, as an SEM does. Its sample requirements and resolution are TEM-like, and point-by-point scanning makes it well suited to EDX and EELS mapping.

How thin must a TEM sample be?

Generally 100 nm or less, and a few tens of nanometres for high-resolution lattice imaging. Higher accelerating voltages penetrate thicker specimens, but thinner samples give sharper images with less overlap of features along the beam direction.

Which Japanese companies make electron microscopes?

Hitachi High-Tech and JEOL are the main Japanese makers of both SEMs and TEMs, and Hitachi High-Tech also supplies CD-SEMs for semiconductor lines. Thermo Fisher Scientific and Carl Zeiss are major non-Japanese suppliers.

Terms compared


Japanese companies involved


More inspection and metrology comparisons


Sources


  1. Toray Research Center: Scanning Electron Microscope (SEM) (Japanese)
  2. Hitachi High-Tech: What is a CD-SEM (Japanese)
  3. Microscopy Society of Japan: TEM lamella preparation with a focused ion beam, Kenbikyo 54(3) (Japanese)(PDF)

More on this topic


Semiconductor comparisons

Inspection and metrology terms