Comparison · Inspection and metrology

XPS vs EDX: what is the difference?

Both are workhorse techniques in semiconductor analysis labs, and their names both contain 'X-ray', which causes confusion. The difference lies in what goes in, what comes out and how deep the signal originates, and those three facts decide which questions each method can answer.

The short answer


XPS (X-ray photoelectron spectroscopy) irradiates a sample with X-rays and measures the kinetic energy of the emitted photoelectrons, which identifies elements in the top few nanometres and their chemical bonding state, such as whether silicon is oxidised. EDX (energy-dispersive X-ray spectroscopy) measures the characteristic X-rays produced when an electron beam hits the sample inside an SEM or TEM, giving the elements present at a chosen spot and maps of their distribution, but almost no chemical-state information. Use XPS for surface and interface chemistry, EDX for identifying particles and mapping cross-sections.

Side-by-side comparison


AspectXPSEDX
ProbeSoft X-raysElectron beam
Signal measuredKinetic energy of emitted photoelectronsEnergy of characteristic X-rays
Information depthTop few nanometresRoughly 1 µm in an SEM; the lamella thickness in a TEM
Chemical stateYes, from chemical shifts in binding energyEssentially no
Lateral resolutionAnalysis areas from tens of micrometres to millimetresThe spot seen in the SEM or TEM image; nanometre scale in a TEM
InstrumentDedicated ultra-high-vacuum spectrometerA detector fitted to an SEM or TEM
Elements not detectedHydrogen and heliumHydrogen and helium; light elements are hard to quantify
Typical detection limitAround 0.1 at.%Around 0.1 wt.%
Typical semiconductor usesInterface bonding, surface state after cleaning or etching, thin-film compositionIdentifying particles and residues, element maps of cross-sections

How XPS reads chemical states


In XPS, an X-ray photon transfers its energy to a core-level electron. The electron leaves the atom with a kinetic energy equal to the photon energy minus its binding energy (and a small instrument work function). Each element has characteristic binding energies, and those energies shift slightly depending on what the atom is bonded to. This chemical shift lets XPS distinguish elemental silicon from silicon dioxide, or a metal from its oxide or nitride.

Photoelectrons lose energy quickly in a solid, so only those from the top few nanometres escape without losses. That makes XPS extremely surface-sensitive, which is ideal for native oxides, cleaning residues and the first layers of a film stack, but it also means the sample needs ultra-high vacuum and careful handling. Depth profiles are built by alternating measurement with ion sputtering, or by varying the take-off angle.

How EDX maps elements


In EDX, the electron beam of an SEM or TEM knocks inner-shell electrons out of atoms. When outer electrons fill the vacancy, the atom emits an X-ray whose energy is characteristic of the element. A detector sorts these X-rays by energy to produce a spectrum.

Because the beam is the same one that forms the image, an analyst can point at a particle or a layer in a cross-section and analyse it immediately, or scan the beam to produce colour-coded maps of each element. The method is fast and convenient. Its limits are the large interaction volume in bulk samples, which blurs results for features smaller than about a micrometre in an SEM, weak sensitivity to light elements, overlapping peaks for some element pairs and the absence of useful chemical-state information.

How they are combined in failure analysis


To identify a particle found on a wafer, the usual first step is SEM imaging with EDX to learn which elements it contains, which points to its likely source in the fab. To understand bonding at a gate oxide interface, or what an etch or clean has left on the surface, XPS is the better tool.

Neither method measures trace impurities well. Dopant concentrations and contamination at ppm levels and below, and their depth profiles, are measured with SIMS (secondary ion mass spectrometry). Labs routinely use all three, choosing by the question rather than by habit.

FAQ


Which should I use, XPS or EDX?

Choose by the information needed. If the question is whether a surface element is oxidised, or what it is bonded to, use XPS. If the question is which elements are present at a specific spot seen in an electron microscope, or how they are distributed in a cross-section, use EDX.

Is EDS the same as EDX?

Yes. EDS, EDX and EDXS are all names for energy-dispersive X-ray spectroscopy. A related method, WDS (wavelength-dispersive spectroscopy), resolves X-ray energies more finely and helps where EDX peaks overlap.

Can either method measure dopant levels?

Not at typical doping concentrations. Both have detection limits around 0.1 per cent, far above most dopant levels. SIMS, which detects ions sputtered from the sample, reaches ppm and lower and gives depth profiles.

Terms compared


Japanese companies involved


More inspection and metrology comparisons


Sources


  1. Toray Research Center: X-ray Photoelectron Spectroscopy (XPS) (Japanese)
  2. JEOL: Glossary, energy-dispersive X-ray spectroscopy (EDS)
  3. Japan Society of Applied Physics: Tips for elemental analysis, Oyo Butsuri 85(4) (Japanese)(PDF)
  4. Toray Research Center: Secondary Ion Mass Spectrometry (SIMS) (Japanese)

More on this topic


Semiconductor comparisons

Inspection and metrology terms

Last updated: November 2, 2026